Leica-biosystems IPS Instrukcja Użytkownika Strona 66

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 65
64
Instructions for Use V 1.9 RevD - 05/2013
7. Troubleshooting
Malfunction / failure
• Instrumentnotfunctioning.
• Nodisplayindication.
• Instrumentnotworkingatnormalspeed.
Printing a slide takes approx. 8 sec., even
after the warm-up phase has been com-
pleted.
• Checkwhetherthereisageneralpowerfailure(nopower).
• Checkwhetherthepowerplugisinsertedcorrectlyintothewalloutlet
and whether the wall outlet is switched on, if applicable.
• Checkwhetherthepowerswitchisswitchedoncorrectly.Possibly,the
primary fuse has tripped. If so, the main switch will be in "0" = OFF posi-
tion (Fig. 63).
• Checkwhetheroneofthetwosecondaryfusesisdefective.(F1, F2
located at the back panel of instrument, Fig. 64, see also Fig. 65).
Some instrument malfunctions / failures are caused by defective fuses.
Fuse F1
Fuse F2
7.5 Power failure
Fuse to be checked
7.4 Changing the flash bulb
Code 13 is displayed when the flash bulb has reached its maximum service-
able life.
When code 13 is issued, the bulb has to be changed.
For details on how to insert / replace the flash bulb, please refer to Chap-
ter 4.7.
Fig. 64
Fig. 63
10 s
Before replacing the flash bulb, switch off the printer and unplug
the instrument from power supply.
After changing the bulb, switch the printer back on.
Then go to Off-line mode and press and hold the ERROR button for 10 sec-
onds - code 13 disappears from the display.
ð
Przeglądanie stron 65
1 2 ... 61 62 63 64 65 66 67 68 69 70

Komentarze do niniejszej Instrukcji

Brak uwag