6. Cleaning and Maintenance
Once a week or if message "15" is
displayed, the print head must be
cleaned manually.
Preparing the printer
• Open theprinter lidandthen
press buttons CLEAN and
LOADED simultaneously.
The print head (72) moves
upward to a position approx.
1 cm away from the sealing lip
(Fig. 53).
6.2 Clean print head
• Carefully inserttheswab (89) into the gap
under the print head and apply light pres-
sure upwards (on print head) and move the
foam swab back and forth (Fig. 55) (approx.
10 times!).
This procedure removes dried ink residues.
Fig. 53
69
70
+
• Pushthe lever(69) upwards, then remove
the red location plate (70) with the sealing lip
(Fig. 54).
• Moistenoneofthefoamswabssuppliedwith
the instrument (73) with some alcohol. Be sure
not to use too much alcohol—no alcohol may
drip into the instrument.
Fig. 54
Fig. 55
72
89
Never use acetone or xylene! Only use
alcohol 95 % or 100 % for cleaning pur-
poses.
Never rotate the swab—this can dam-
age the nozzle plate of the print head.