6. Cleaning and Maintenance
6.4 Storing the printer
If the printer is not used for a long period of time (up to three months), the
following instructions are to be followed.
Important!
• If the instrument remains connected to the power supply with the power
on, the ink cartridge can remain in the printer up to its expiration date.
• The print head drying out is prevented, because the printer regularly
rinses the print head.
• Youmustabsolutelyensurethattheinkcartridgeisswitchedoutaccord-
ing to its expiration date.
If the instrument is switched off and separated from the power supply, the
following is to be done:
• Openthe coverplate (5) on the left side of
the instrument by pressing its top left corner
(see Fig. 31). Then push the security latch (86)
forward (Fig. 57).
• Pull theinkcartridge (66) approx. 3 cm out
of the instrument so that the "Ink Empty" LED
lights.
• Screw theredcap (87) to the shutter (88),
close it securely and take the ink cartridge
out completely.
• Storetheinkcartridgeinahorizontalposition
in a sealed container.
Instructions on the storage location
Fig. 57
88
86
87
66
A printer may be stored for a maximum of three months after fol-
lowing the procedure described below. Storing the printer for lon-
ger periods may result in damage to the print head.
The ink cartridge cannot be used in
another printer, because the ink level
information is stored in the printer
itself.